Table 2. Advantages and limitations of MS-based methods and immunoassays for vitamin D measurement.

Method Advantages Limitations
MS-based high sample volume (LC-MS)
availability of reference measurement procedures
separation of vitamin D metabolites
allows the inclusion of stable isotope-labelled standards
not intended for routine clinical samples analysis
not fully automated
complex and expensive instrumentation
no distinction between epimers
Immunoassays high sensitivity
reduced sample volume
fully automated
cross-reactivity with other vitamin D metabolites (e.g., D2 vs. D3)
measurement of free 25(OH)D requires VDBP displacement (proprietary methods)
LC-MS – liquid chromatography-mass spectrometry. VDBP - vitamin D binding proteins.