Skip to main content
. Author manuscript; available in PMC: 2018 Feb 1.
Published in final edited form as: Anal Bioanal Chem. 2016 Nov 12;409(5):1323–1332. doi: 10.1007/s00216-016-0061-4

Table 4.

The mean slope of calibration curve ± standard deviation in water and in urine for each flame retardant metabolite, and % difference between the slopes

Analyte Slope ± standard deviationa
% Difference
Water curve Urine curve
BCEtP 0.1048 ± 0.0034 0.1058 ± 0.0022 1.0
BCPP 0.0921 ± 0.0005 0.0894 ± 0.0025 2.9
BDCPP 0.1001 ± 0.0013 0.0956 ± 0.0027 4.6
DBuP 0.1130 ± 0.0008 0.1173 ± 0.0073 3.7
DBzP 0.1065 ± 0.0026 0.1048 ± 0.0040 1.7
DPhP 0.1185 ± 0.0019 0.1198 ± 0.0097 1.0
DoCP 0.0912 ± 0.0008 0.0870 ± 0.0018 4.7
DpCP 0.0986 ± 0.0031 0.0942 ± 0.0021 4.6
TBBA 0.0806 ± 0.0016 0.0780 ± 0.0017 3.2
a

N = 10