Matrigel solidifies in pipette tip |
Matrigel has reached room temperature |
Work quickly |
Mattress in oblique line |
Pipetting too quickly |
Slow down |
Poor quality mattresses |
Preparing too many mattress simultaneously |
Prepare less until proper technique is established |
Mattress implodes |
Short incubation |
Incubate mattresses for 8–10 minutes prior to seeding hiPSC-CMs |
hiPSC-CMs do not contract on mattress |
Mattresses are too thick or thin |
Apply Matrigel at a thickness of 0.4 mm |
Cells dissociate from mattress |
Perfusion or aspiration too strong during daily media changes |
Decrease aspiration strength |
Mattress ruptures or dislodges from dish |
Media forcefully applied to dish |
Gently add media to the corner of the dish; not directly on the mattress |