Table 2. Error quantification of AView measurement using anatomical models of IAs with known sizes.
IA Models | Size (mm) | Volume (mm3) | ||||
---|---|---|---|---|---|---|
AView | True Value | Error | AView | True Value | Error | |
Giant ICA | 20.77 | 21.27 | -2.35 % | 4291.99 | 4304.38 | -0.29 % |
ACA | 14.3 | 14.42 | -0.83 % | 706.76 | 700.81 | 0.85 % |
ICA | 9.07 | 8.99 | 0.89 % | 226.35 | 227.41 | -0.47 % |
BA | 5.04 | 4.84 | 4.13 % | 100.17 | 95.07 | 5.36 % |
MCA | 4.91 | 4.91 | 0.00 % | 87.49 | 89.37 | -2.10 % |
Blister ICA | 1.51 | 1.56 | -3.21 % | 3.92 | 3.82 | 2.62 % |