Table 1.

Reproducibility of AutoTip on analysis N-glycans from Bovine serum.

Composition Structure Isomer AutoTip Set Mean
1 2 3
RT (min) Intensity CV (%) RT (min) Intensity CV (%) RT (min) Intensity CV (%) RT (min) CV (%) Intensity
S1H5N4 graphic file with name 41598_2017_10487_Figa_HTML.gif A 17 2404 16.8 16.3 2104 7.7 16.08 1209 11.8 16.5 12.1 1906
B 25.8 49401 7.3 25.17 26239 7 24.94 38961 6.8 25.3 7 38200
S2H5N4 graphic file with name 41598_2017_10487_Figb_HTML.gif A 25.8 83909 8.9 25.23 17578 11.2 25.01 55815 11.4 25.4 10.5 52434
S2H6N5 graphic file with name 41598_2017_10487_Figc_HTML.gif A 25.4 31505 11.8 24.58 28297 13.5 24.64 31891 10.9 24.9 12.1 30564
S3H6N5 graphic file with name 41598_2017_10487_Figd_HTML.gif A 32.5 193155 10.7 30.51 159417 9.4 31.34 140108 11.1 31.5 10.4 164227
B 67.9 12742 15.2 80.76 1305 18.6 80.3 1136 16.3 76.3 16.7 5061
C 109.6 1404 20.8 10.7.27 3461 15.3 107.79 3141 17.1 108.7 17.7 2669
S4H6N5 graphic file with name 41598_2017_10487_Fige_HTML.gif A 48.2 21628 19.4 44.85 8292 16.8 45.53 13784 18.4 46.2 18.2 14568
B 64.6 31540 18 61.33 9396 17.7 60.27 8926 15.9 62.1 17.2 16621

N-glycans were released from AutoTips using an automated liquid handler. The reproducibility was calculated by quantitative analysis of N-glycans from four independent AutoTip by 4-plex QUANTITY tags (176, 177, 178, and 179). The labeled N-glycans were measured by LC-ESI-MS/MS (Thermo Orbitrap Velos). Experiments were performed in triplicates. CV (coefficient of variation) was average of the triplicated tests.