. 2017 Apr 3;27(10):4351–4359. doi: 10.1007/s00330-017-4801-4

Table 2.

Mean measurement errors. Subdivided by nominal design concentrations, mAs and kVp

True BMD concentration (mg/cm3) mAs 120 kVp 140 kVp
Measurement error Measurement error
mg/cm3 % mg/cm3 %
50* All -0.9 ± 2.1 -1.8 ± 4.0 1.6 ± 1.1 3.1 ± 2.1
200 0.3 ± 1.5 0.5 ± 2.9 0.9 ± 1.0 1.8 ± 1.9
100 -0.2 ± 0.6 -0.5 ± 1.1 1.5 ± 0.6 2.9 ± 1.1
50 -2.8 ± 2.6 -5.5 ± 5.1 2.3 ± 1.5 4.5 ± 2.9
100* All -3.0 ± 2.7 -2.9 ± 2.6 -0.7 ± 1.8 -0.7 ± 1.7
200 -2.7 ± 1.8 -2.6 ± 1.8 -0.5 ± 2.0 -0.5 ± 1.9
100 -2.1 ± 3.8 -2.0 ± 3.7 -0.5 ± 2.0 -0.5 ± 1.9
50 -4.1 ± 2.2 -4.0 ± 2.1 -1.1 ± 1.6 -1.1 ± 1.5
200* All -10.6 ± 5.3 -5.3 ± 2.7 -6.0 ± 2.7 -2.9 ± 1.3
200 -9.6 ± 4.6 -4.7 ± 2.3 -6.4 ± 1.6 -3.2 ± 0.8
100 -10.5 ± 5.2 -5.2 ± 2.6 -6.7 ± 3.1 -3.3 ± 1.5
50 -11.8 ± 6.7 -5.8 ± 3.4 -4.8 ± 3.1 -2.4 ± 1.6

Data are given as mean ± standard deviation. For 50 mg/cm3 nine measurements and for 100 and 200 mg/cm3 18 measurement were performed at both 120 and 140 kVp (three and six,respectively, per mAs value)

BMD bone mineral density, kVp kilovoltage peak, mAs milliampere second

* Phantom specific concentrations (ESP-143; 51.2, 102.3 and 201.2 mg/cm3 HA and BDC-03-29; 104.4 and 206.2 mg/cm3 HA) were used to determine measurement error