Comparison of Cr-SPE, HMDE and AAS for the determination of Zn(II), Cd(II), Pb(II) and Cu(II) in real wastewater samples.

Type of Measurement Zn(II) (mg∙L−1) Cd(II) (mg∙L−1) Pb(II) (mg∙L−1) Cu(II) (mg∙L−1)
Cr-SPE 2.6 ± 0.8 3.5 ± 0.7 5.7 ± 1.1 7.9 ± 0.6
HMDE 1 6.2 ± 0.5 3.9 ± 0.3 4.7 ± 0.3 8.3 ± 0.5
AAS 6.6 ± 0.01 4.2 ± 0.04 4.8 ± 0.02 8.8 ± 0.01