Skip to main content
. Author manuscript; available in PMC: 2018 Oct 1.
Published in final edited form as: Ultrasound Med Biol. 2017 Jul 14;43(10):2302–2317. doi: 10.1016/j.ultrasmedbio.2017.06.010

Table 2. −6-dB Widths of the Pressure Profiles Measured through the Skullcaps.

Table 1 shows the −6-dB widths of the 1D pressure profiles measured through the three skullcaps.

Skull # Sagittal (mm) Coronal (mm) Axial (mm)
250 kHz 500 kHz 250 kHz 500 kHz 250 kHz 500 kHz
1 3.7 2.0 3.2 1.8 7.6 3.8
2 3.7 2.0 3.2 1.7 7.0 3.6
3 3.4 1.9 3.2 1.8 6.9 3.5