Skip to main content
. Author manuscript; available in PMC: 2017 Sep 1.
Published in final edited form as: Aerosol Sci Technol. 2016 Mar 10;50(5):462–473. doi: 10.1080/02786826.2016.1162901

Table 4.

Regression equations to estimate mass concentration (y in μg/m3) from DC1700 number concentration (x = small bin particles/cm3). For particle concentrations (x) less than 106 particles/cm3

Aerosol Equation R2
0.9% Salt Aerosol y = 2.6 × x − 22 0.96
5% Salt Aerosol y = 6.2 × x − 153 0.99
Arizona Road Dust y = 8.6 × x + 98 0.98
Diesel Fume y = 54 × x + 511 0.91
Welding Fume y = 12 × x − 12 0.96