Table 5.
Evaluation of mass concentrations measured with the pDR-1500 with reference to those measured with the SMPS and APS.
Aerosol | Data Pairs | Slope ± Std. Error | Intercept ± Std. Error (μg/m3) | r | R2 | % Bias |
---|---|---|---|---|---|---|
0.9% Salt Aerosol | 21 | 1.2 ± 0.02 | −46 ± 12 | 0.991 | 0.99 | −3.41,2 |
5% Salt Aerosol | 21 | 0.9 ± 0.0061,2 | −5.0 ± 4.01 | 0.991 | 0.99 | 4.61,2 |
Arizona Road Dust | 18 | 1.3 ± 0.04 | −54 ± 14 | 0.991 | 0.98 | −2.01,2 |
Diesel Fume | 21 | 1.1 ± 0.021,2 | 28 ± 66 | 0.991 | 0.99 | −9.11,2 |
Welding Fume | 18 | 1.0 ± 0.021,2 | −55 ± 24 | 0.991 | 0.99 | −1.11,2 |
meets EPA criterion