Gas chromatography–mass spectrometry conditions for VOC analysis

GC
 Instrument 7890 (Agilent Technologies, Santa Clara, CA)
 Column SUPELCOWAX 10, 60 m × 0.32 mm, 0.5 μm (Sigma-Aldrich)
 Injection method Split (1:450)
 Injection volume 1.0 μL
 Carrier gas Helium
 Flow rate 1.0 mL/min
 Injector temperature 280 °C
 Oven temperature program 60 °C (held 5 min) → 230 °C at 6 °C/min (held 2.67 min)
 Transfer line temperature 230 °C
MS
 Instrument Quattro micro (Waters, Milford, MA)
 Mode Selected ion monitoring
 Ion source temperature 230 °C
 Ionization mode Electron ionization