Table 2. Validation of lyophilized assays.
Genomic E. coli DNA Assay | High DNA Template | Medium DNA Template | Low DNA Template |
Control | 15.71 +/- 0.16 | 23.55 +/- 1.5 | 29.14 +/- 0.15 |
Lyophilized | 16.43 +/- 0.51 | 22.87 +/- 0.43 | 29.87 +/- 0.32 |
E. coli RNA Assay | dnaK-FAM Ct | rpoA-HEX Ct | srlR-Cy5 Ct |
Control | 25.79 +/- 0.31 | 24.4 +/- 0.11 | 29.32 +/- 0.33 |
Lyophilized | 24.99 +/- 1.52 | 23.27 +/-0.16 | 28.53 +/- 0.56 |
Mouse Liver RNA Assay | gapdh-FAM Ct | rpl19-HEX Ct | fn1-Cy5 Ct |
Control | 21.11 +/- 0.085 | 19.35 +/- 0.092 | 21.57 +/- 0.092 |
Lyophilized | 21.59 +/- 0.41 | 19.33 +/- 0.33 | 21.17 +/- 0.085 |
Lyophilized and non-lyophilized assay comparison of qPCR cycle threshold data for E. coli DNA Assay for (High, Mid, Low) DNA template amount, and RT-qPCR cycle threshold data for E. coli and mouse liver RNA. There are no statistical differences between lyophilized assays and commercial non-lyophilized assays. Data shows Ct values +/- SD.