Table 3.

Area under curve (AUC) of Corneal Visualization Scheimpflug Technology (Corvis ST) parameters with their standard error (SE), 95% confidence interval (CI), and their best cut-offs for optimizing sensitivity and specificity to separate normal and keratoconic corneas.

Variable Cut-off point Sensitivity (%) Specificity (%) AUC SEa 95% CIb
A1T (ms) ≤7.03 93.75 92.00 0.955 0.0218 0.893–0.986
A1L (mm) ≤1.67 50.00 94.00 0.675 0.0594 0.573–0.766
A1V (m/s) >0.12 68.75 64.00 0.692 0.0533 0.591–0.782
A2T (ms) >21.4 68.75 98.00 0.898 0.0309 0.805–0.941
A2V (m/s) ≤−0.37 83.33 82.00 0.862 0.0385 0.777–0.923
HCDA (mm) >1 72.9 94.00 0.893 0.0328 0.814–0.946
HCR (mm) ≤6.35 77.08 100.00 0.936 0.0242 0.868–0.976

AUC: Area under curve; SE: Standard error; CI: Confidence interval; A1T: First applanation time; A1L: First applanation length; A1V: First applanation velocity; A2T: Second applanation time; A2V: Second applanation velocity; HCDA: Highest concavity deformation amplitude; HCR: Highest concavity radius.

a

DeLong et al., 1988.

b

Binomial exact.