Table 5.

Area under curve (AUC) of Corneal Visualization Scheimpflug Technology (Corvis ST) parameters with their standard error (SE), 95% confidence interval (CI), and their best cut-offs for optimizing sensitivity and specificity to separate normal and keratoconic corneas in two central corneal thickness (CCT) controlled subgroups.

Variable Cut-off point Sensitivity (%) Specificity (%) AUC SEa 95% CIb
A1T (ms) ≤7 90 90 0.904 0.0545 0.768–0.974
A2T (ms) >21.4 70 100 0.879 0.0538 0.737–0.960
A2V (m/s) ≤−0.37 75 75 0.737 0.0822 0.575–0.864
HCDA (mm) >1.1 55 100 0.794 0.0720 0.636–0.905
HCR (mm) ≤6.13 65 100 0.821 0.0703 0.668–0.924

AUC: Area under curve; SE: Standard error; CI: Confidence interval; A1T: First applanation time; A2T: Second applanation time; A2V: Second applanation velocity; HCDA: Highest concavity deformation amplitude; HCR: Highest concavity radius.

a

DeLong et al., 1988.

b

Binomial exact.