Skip to main content
. Author manuscript; available in PMC: 2018 Oct 25.
Published in final edited form as: J Pharm Biomed Anal. 2017 Jul 31;145:629–633. doi: 10.1016/j.jpba.2017.07.049

Table 1.

Intraday (n=5) and interday (n=15, three runs) accuracy and precision results in mouse plasma

QC Concentration (μM) Intraday (n=5) Interday (n=15, three runs)

Conc. CV (%) Accuracy (%) Conc. CV (%) Accuracy (%)
LLOQ (0.025) 0.0259 11 104 0.0238 4.9 95
LQC (0.100) 0.100 2.3 100 0.0983 1.3 98
MQC (1.00) 1.05 1.9 105 1.02 1.9 102
HQC (25.0) 23.9 4.0 95 23.7 2.3 95