Skip to main content
. 2017 Sep 6;7:7499. doi: 10.1038/s41598-017-06975-6

Table 2.

Mean PaO2 value, oscillations amplitudes and mean airway pressure associated with different ventilatory modes (volume or pressure), I:E ratios ranging from 1:4 to 4:1, tidal volume of 10 ml kg−1, and respiratory rate of 12 breaths per minute.

I:E Volume control Pressure control
PaO2 mean (mmHg) PaO2 oscillation amplitude (mmHg) Airway Pressure (cmH2O) PaO2 mean (mmHg) PaO2 oscillation amplitude (mmHg) Airway Pressure (cmH2O)
1:2 127 ± 12 10 ± 5 5 ± 2 134 ± 14 10 ± 6 6 ± 3
2:1 127 ± 14 7 ± 4 6 ± 2 140 ± 13 8 ± 4 8 ± 3
1:4 125 ± 10 10 ± 5 5 ± 2 133 ± 9 10 ± 4 5 ± 3
4:1 129 ± 12 5 ± 3 7 ± 2 144 ± 8 7 ± 3 9 ± 2

Values are mean ± standard deviation recorded in 6 animals for 2 min at steady state. RR: respiratory rate. VT: tidal volume.