Skip to main content
. 2017 Aug 16;37(6):494–498. doi: 10.3343/alm.2017.37.6.494

Table 2. Analytical sensitivity of the PowerChek MERS (upE and ORF1a) real-time PCR Kit.

Concentration (copies/µL) N of detections/N of replicates (%)
upE ORF1a
100 20/20 (100) 20/20 (100)
50 20/20 (100) 20/20 (100)
25 20/20 (100) 20/20 (100)
12.5 17/20 (85) 20/20 (100)
10 18/20 (90) 19/20 (95)