Table 3. Precision of the PowerChek MERS (upE and ORF1a) real-time PCR Kit.

Target Concentration (copies/µL) Mean Ct (% coefficient of variation)
Reproducibility Repeatability Lot-to-lot variation
Site 1 Site 2 Lot 1 Lot 2 Lot 3
upE 50 34.40 (2.10) 34.40 (1.32) 34.74 (2.11) 34.40 (2.10) 34.45 (3.36) 34.40 (2.10)
150 32.68 (0.89) 32.85 (1.19) 32.92 (1.06) 32.68 (0.89) 32.82 (0.97) 32.68 (0.89)
300 31.91 (0.54) 31.99 (0.88) 31.97 (0.77) 31.91 (0.54) 31.98 (0.68) 31.91 (0.54)
ORF1a 50 33.93 (1.30) 33.90 (0.86) 34.07 (1.53) 33.93 (1.30) 34.14 (1.34) 34.04 (1.21)
150 32.16 (0.79) 32.11 (0.63) 32.23 (0.77) 32.16 (0.79) 32.31 (0.77) 32.27 (0.49)
300 30.88 (0.62) 30.78 (0.62) 30.91 (0.54) 30.88 (0.62) 30.85 (1.15) 30.98 (0.70)

Abbreviation: Ct, Cycle threshold.