Table 2. Validity results.

Overall Bias Limits of Agreements
Absolute Relative (%) Absolute Relative (%) Pearson CC CV (%)
Volume 30.4 (28.3; 32.5) ml 1.40 (1.28; 1.51) 18.9 ml 1.01 0.99 0.34 (0.29; 0.40)
Residuum Size
Width (x) 0.49 (0.39; 0.59) mm 0.36 (0.29; 0.43) 0.91 mm 1.01 0.99 0.32 (0.28; 0.38)
Depth (y) 0.47 (0.38; 0.55) mm 0.35 (0.29; 0.41) 0.77 mm 1.01 0.99 0.24 (0.21; 0.28)
Length (z) 0.38 (0.26; 0.50) mm 0.14 (0.10; 0.19) 1.09 mm 1.00 0.99 0.20 (0.17; 0.23)
BCOM
x -0.01 (-0.04; 0.03) mm - 0.32 mm - 0.99 -
y 0.02 (-0.04; 0.07) mm - 0.48 mm - 0.99 -
z 0.23 (0.05; 0.41) mm - 1.63 mm - 0.99 -

Validity results are presented in terms of overall bias (absolute and relative values), limits of agreements, Pearson Correlation Coefficient (CC) and Coefficient of Variation (CV). Values in brackets represent the 95% Confidence Limits.