. 2017 Sep 11;7:11238. doi: 10.1038/s41598-017-11680-5

Table 2.

Values of the PSA concentrations in the patients’ blood serum obtained by the aptasensor.

Nominal valuea (ng mL−1) Obtained value (ng mL−1) Bias (%)
5.31 5.2 −2.1
6.78 6.94 2.4
6.87 6.99 1.7
8.48 8.54 0.7
8.71 8.72 0.1
9.19 8.95 −2.6
10.92 11.01 0.8
12.31 11.88 −3.5
13.31 13.60 2.2
19.00 18.46 −2.8

aNominal values were obtained by the immunoradiometric assay.