Table 1.

Limits of detection (LOD) and within-run coefficient of variation (CV) from 10 LOD experiments measured over a period of about one year.

Experiment no. Date of measurement Cartridge lot Instrument no. LOD [pg/mL] Within-run CV at
+30 pg/ml BNP equivalents (n=5) +60 pg/ml BNP equivalents (n=5) +90 pg/ml BNP equivalents (n=5)
1 28 Jan 2013 A 1 1.9 7.6% 3.4% 5.7%
2 07 Feb 2013 B 1 1.9 3.9% 2.7% 4.4%
3 22 Mar 2013 C 1 1.9 4.4% 4.5% 4.2%
4 02 Mai 2013 D 2 2.5 2.6% 11.5% 3.4%
5 17 Jun 2013 E 3 4 6.1% 5.2% 5.9%
6 18 Jun 2013 F 3 4.3 5.0% 2.3% 6.5%
7 03 Sep 2013 G 2 1.8 6.0% 4.4% 3.6%
8 30 Oct 2013 H 4 1.6 4.0% 5.8% 4.8%
9 13 Feb 2014 I 1 1.4 4.3% 2.9% 4.5%
10 13 Feb 2014 I 5 1.8 3.7% 3.1% 1.9%
Mean value±standard deviation: 2.3±1.0 4.8±1.4% 4.6±2.7% 4.5±1.4%