Table 1. Comparison between the three different protocols of EVLP.

Parameters Lund Toronto OCS
Perfusion
Target flow 40% of CO 40% of CO 2–2.5 L/min
Pressure
   PA Flow dictated ≤20 mmHg ≤20 mmHg
   LA 3–5 mmHg (closed) 0 mmHg (open) 0 mmHg (open)
Perfusate Steen TM solution Steen TM solution and RBCP Hct 14% OCS TM solution and PRBC Hct 15–25%
Pump Roller Centrifugal Piston (pulsatile)
Ventilation
   Start temp (°C) 32 32 34
   Tidal volume 7 mL/kg 5–7 mL/kg 7 mL/kg
   RR 7 20 10
   PEEP 5 cmH2O 5 cmH2O 5–7 cmH2O
   FiO2 (%) 21 50 12
   Sweep gas flow Titrate to PA pCO2 34–38 mmHg Titrate to PA pCo2 34–38 mmHg
Temperature (°C)
   Start of ventilation 32 32 32
   Start of perfusion 15 25 32
   Start of evaluation 37 37 37

EVLP, ex vivo lung perfusion; OCS, Organ Care System; CO, cardiac output; PA, pulmonary artery; LA, left atrial; RBCP, red blood cell products; Hct, hematocrit; RR, respiratory rate; PEEP, positive end expiratory pressure; FiO2, fraction of inspired oxygen.