Skip to main content
. 2017 Jul 6;18(5):29–35. doi: 10.1002/acm2.12130

Table 1.

Dose‐volume constraints

Structure Dose volume constraints (Gy[RBE]) Achieved values from MIP (Gy[RBE]) Achieved values from L‐BFGS (Gy[RBE])
Prostate (with 3837 beamlets)
Planning target volume D95% >44 44 44
Planning target volume D5%–D95% <2 1.6 2
Bladder D25% <45 44.2 44.5
Femoral heads D10% <28 12.4 14.1
Rectum D25% <45 45 44.8
H&N, case 1 (with 7773 beamlets)
Planning target volume D98% >59 60 59.3
Planning target volume D2%–D98% <2 0 1.3
Whole brain D1% <62 60 59.5
Brain stem D1% <55 55 53
Optic chiasm D1% <55 55 56.1
H&N, case 2 (with 9905 beamlets)
Planning target volume D98% >67 68 67.6
Planning target volume D2%–D98% <2 0 0.8
Brain stem D1% <55 25.2 24.3
Mucosa avoid D1% <30 15.4 16.3
Spinal cord D1% <45 20.2 21.9