Table 1.

Lower and upper limits of quantitation (LLOQ and ULOQ) and extraction recovery. For the LLOQ, the accuracy had to be within 80–120% with less than 20% imprecision and the signal had to be at least five-fold higher than that of a blank sample. Recovery was determined in samples collected from 6 different individuals at 3 different levels for each analyte. Mean recoveries ± standard deviation (SD) are shown.

Plasma Urine

LLOQ ULOQ % Recovery LLOQ ULOQ % Recovery
ng/mL ng/mL mean SD ng/mL ng/mL mean SD
11OH-THC 1.56 400 103 3.6 1.56 400 97 5.1
CBN 1.56 400 73 3.4 1.56 400 96 4.4
CBG 0.78 400 86 1.4 0.78 400 96 1.8
THC 0.39 400 74 1.1 0.78 400 95 3.6
THCV 0.78 400 93 3.3 0.78 400 98 4.7
THC-COOH 0.39 400 73 3.1 1.56 400 100 2.1
THCV-COOH 0.78 400 88 3.4 3.13 400 101 4.5
CBD 0.39 400 87 3.6 0.78 400 98 4.7
CBC 0.78 400 60 1.3 0.78 400 94 3.0
CBDV 0.78 400 98 2.4 0.78 400 99 3.3
THC-C-gluc 3.91 2000 71 6.0 3.91 2000 99 3.2