Table 2.
Mean Hounsfield unit (HU) values and objective image noise (SD) ± standard deviation of standard of care and reduced dose CT series (0.25, 0.4 and 0.8 mGy) for iterative reconstruction techniques [SafeCT (MedicVision®, Tirat Carmel, Israel), adaptive statistical iterative reconstruction (ASIR; GE Healthcare, Waukesha, WI) and model-based iterative reconstruction (MBIR; GE Healthcare)]
| ASIR40 | ASIR70 | ASIR90 | SafeCT-CH0 | SafeCT-CH1 | MBIR | p-value | ||
|---|---|---|---|---|---|---|---|---|
| Descending thoracic aorta | ||||||||
| Standard of care CT | HU | 206 ± 97 | ||||||
| SD | 16 ± 5 | |||||||
| 0.25 mGy | HU | 147 ± 92 | 146 ± 91 | 146 ± 91 | 144 ± 91 | 55 ± 101 | 0.003 | |
| SD | 55 ± 13 | 49 ± 13 | 59 ± 13 | 48 ± 12 | 40 ± 15 | <0.001 | ||
| 0.4 mGy | HU | 148 ± 58 | 148 ± 58 | 147 ± 58 | 147 ± 57 | 138 ± 58 | 0.9 | |
| SD | 40 ± 10 | 34 ± 9 | 43 ± 11 | 34 ± 9 | 18 ± 7 | <0.001 | ||
| 0.8 mGy | HU | 131 ± 52 | 131 ± 52 | 131 ± 52 | 130 ± 52 | 130 ± 53 | 0.9 | |
| SD | 27 ± 9 | 22 ± 8 | 29 ± 10 | 21 ± 8 | 16 ± 4 | <0.001 | ||
| Tracheal air | ||||||||
| Standard of care CT | HU | −938 ± 21 | ||||||
| SD | 25 ± 13 | |||||||
| 0.25 mGy | HU | −814 ± 123 | −814 ± 124 | −815 ± 125 | −814 ± 131 | −757 ± 146 | 0.5 | |
| SD | 67 ± 22 | 66 ± 21 | 65 ± 21 | 60 ± 22 | 28 ± 12 | <0.001 | ||
| 0.4 mGy | HU | −933 ± 36 | −933 ± 36 | −934 ± 36 | −935 ± 37 | −916 ± 42 | 0.4 | |
| SD | 52 ± 18 | 51 ± 17 | 47 ± 17 | 40 ± 18 | 23 ± 12 | <0.001 | ||
| 0.8 mGy | HU | −947 ± 22 | −947 ± 22 | −946 ± 23 | −947 ± 22 | −941 ± 23 | 0.8 | |
| SD | 37 ± 12 | 36 ± 12 | 33 ± 15 | 27 ± 13 | 18 ± 12 | <0.001 | ||
CH0 and CH1 are two settings of SafeCT.