Table 2.
Tube voltage (kV) | Tube current (mA) | RD (mSv) | IR | FBP |
---|---|---|---|---|
120 | 10 | 0.171 | 0.743 ± 0.023 | 0.725 ± 0.026 |
120 | 20 | 0.229 | 0.747 ± 0.027 | 0.734 ± 0.026 |
120 | 80 | 0.614 | 0.788 ± 0.025a | 0.739 ± 0.024 |
120 | 160 | 1.142 | 0.798 ± 0.024b | 0.720 ± 0.027 |
100 | 80 | 0.368 | 0.776 ± 0.025c | 0.751 ± 0.026 |
80 | 320 | 0.570 | 0.796 ± 0.025b | 0.761 ± 0.027 |
FBP, filtered back projection; IR, iterative reconstruction; RD, estimated radiation dose.
Significantly higher than the images acquired using IR at 120 kV/10 mA and using FBP at 120 kV/10, 20, 80 and 160 mA.
Significantly higher than the images acquired using IR at 120 kV/10 mA and 20 mA and using FBP at 120 kV/10, 20, 80, and 160 mA, and 100 kV/80 mA.
Significantly higher than the images acquired using FBP at 120 kV/10 and 160 mA.