. 2017 Feb 24;90(1071):20160555. doi: 10.1259/bjr.20160555

Table 2.

Area under the receiver operating characteristic curve for 10 observers

Tube voltage (kV) Tube current (mA) RD (mSv) IR FBP
120 10 0.171 0.743 ± 0.023 0.725 ± 0.026
120 20 0.229 0.747 ± 0.027 0.734 ± 0.026
120 80 0.614 0.788 ± 0.025a 0.739 ± 0.024
120 160 1.142 0.798 ± 0.024b 0.720 ± 0.027
100 80 0.368 0.776 ± 0.025c 0.751 ± 0.026
80 320 0.570 0.796 ± 0.025b 0.761 ± 0.027

FBP, filtered back projection; IR, iterative reconstruction; RD, estimated radiation dose.

a

Significantly higher than the images acquired using IR at 120 kV/10 mA and using FBP at 120 kV/10, 20, 80 and 160 mA.

b

Significantly higher than the images acquired using IR at 120 kV/10 mA and 20 mA and using FBP at 120 kV/10, 20, 80, and 160 mA, and 100 kV/80 mA.

c

Significantly higher than the images acquired using FBP at 120 kV/10 and 160 mA.