Time to detection (TTD) and detection rates from BacT/ALERT® 3D and VIRTUO™ by bottle type and blood volume

Bottle type Blood volume (mL) 3D VIRTUO™ TTD difference (3D - VIRTUO™)
# (+)/# Detection rate (%) # (+)/# Detection rate (%) Mean TTD difference (h) 95% CI for mean TTD difference (h)
FA Plus 0 227/239 95.0 233/233 100.0* 3.62 (3.37, 3.86)
4** 271/272 99.6 275/275 100.0 3.94 (3.47, 4.41)
10 273/278 98.2 279/279 100.0* 3.24 (2.96, 3.51)
Combined 771/789 97.7 787/787 100.0* 3.59 (3.39, 3.80)
FN Plus 0 93/93 100.0 91/91 100.0 3.62 (3.14, 4.10)
4 117/121 96.7 122/122 100.0* 3.78 (3.27, 4.29)
10 110/115 95.7 114/114 100.0* 4.28 (3.93, 4.64)
Combined 320/329 97.3 327/327 100.0* 3.90 (3.64, 4.17)
SA 0 215/215 100.0 213/213 100.0 2.35 (2.13, 2.56)
4 276/276 100.0 279/279 100.0 3.47 (3.33, 3.61)
10 270/270 100.0 276/276 100.0 3.38 (3.14, 3.62)
Combined 761/761 100.0 768/768 100.0 3.14 (3.02, 3.25)
SN 0 125/125 100.0 123/123 100.0 4.65 (4.46, 4.84)
4 124/124 100.0 116/117 99.1 4.02 (3.76, 4.29)
10 119/120 99.2 119/120 99.2 2.20 (1.58, 2.82)
Combined 368/369 99.7 358/360 99.4 3.65 (3.42, 3.88)
Overall 2220/2248 98.8 2240/2242 99.9* 3.48 (3.38, 3.58)

*Significant difference between detection rates; Fisher’s exact test p-value <0.05; **represents PF Plus