Abstract
Objective:
Hypertension is the leading attributable cause of cardiovascular mortality worldwide. Patients with hypertension have multiple comorbidities including high rates of concomitant renal disease. Current pharmacological approaches are inadequate in the treatment of resistant hypertension. Renal sympathetic denervation (RDN) has been shown to effectively treat resistant hypertension. The traditional use of iodinated contrast in RDN is contraindicated in patients with significant renal insufficiency. In patients with renal impairment, carbon dioxide (CO2) can be used as an alternative contrast material for RDN. This article describes the technical aspects of RDN using CO2 angiography.
Methods:
Our centre is experienced in the innovative RDN procedure using CO2 angiography. We describe the protocol for CO2 angiography for RDN using a home-made CO2 delivery system and the Symplicity™ (Minneapolis MN 55432 USA) catheter (Medtronic) device.
Results:
CO2 angiography is an excellent alternative to iodinated contrast for RDN procedures.
Conclusion:
CO2 angiography for RDN is a safe and effective alternative to iodinated contrast. RDN using CO2 angiography is an easy and feasible procedure that can be used in patients with renal insufficiency or iodinated contrast allergies.
Advances in knowledge:
There is a paucity of descriptive reports for CO2 angiography for RDN and we provide details of the optimal protocol for the procedure. In particular, we describe the use of a Symplicity Spyral™ catheter (Medtronic), which has not been reported to date for use in this procedure.
INTRODUCTION
Hypertension poses a huge burden of disease and is estimated to affect 30–40% of the population in the developed world.1 Patients with hypertension have multiple comorbidities including concomitant renal disease. Current pharmacological antihypertensive treatments are ineffective in patients with resistant hypertension. Resistant hypertension is defined as a blood pressure of ≥140/90 mmHg whilst on at least three maximum-dose antihypertensives, including a diuretic.
Renal sympathetic denervation (RDN) has been studied in international, multicentre randomized control trials as a potential treatment for patients with resistant hypertension.2,3 RDN involves catheter-based denervation using radiofrequency ablation under fluoroscopic guidance. Patients with renal insufficiency are at risk of complications from conventional iodinated contrast materials and consequently their treatment options may be limited.
Carbon dioxide (CO2) angiography provides a safe and effective alternative to iodinated contrast in patients with CKD or iodinated contrast agent allergy.4,5 Owing to its low atomic number, CO2 acts by displacing blood to provide a negative contrast image with digital subtraction angiography. The use of CO2 as a contrast agent has potential serious complications such as neurotoxicity and arrhythmias and its use is therefore contraindicated above the diaphragm, in patients with an arteriovenous malformation, fistula or atrial septal defects. Reports of home-made delivery system CO2 angiography in patients with Chronic kidney disease (CKD), from the study by Kawasaki et al,6 show 100% technical success with only transient and minor complications, predominantly limb or abdominal pain. Complications related to CO2 angiography are reduced by appropriate injection techniques.
We have a large centre in the UK with an established experience with the use of CO2 angiography in renal artery intervention. The centre has extensive use of CO2 angiography for peripheral angiography and endovascular aneurysm repair over a 5-year period. Over a 12-month period, 11 patients aged 49–66 years with resistant hypertension have undergone renal artery denervation using CO2 angiography performed by 2 consultant interventional radiologists. All of these procedures were technically successful and long-term therapeutic effects are currently being investigated. This article describes the technical aspects of performing RDN using CO2 angiography and in particular the use of a second-generation multielectrode Symplicity Spyral™ catheter (Medtronic) for CO2 angiography in RDN.
IMAGING TECHNIQUE: CARBON DIOXIDE ANGIOGRAPHY AND RENAL SYMPATHETIC DENERVATION
CO2 requires specialist equipment to eliminate air contamination and excess compression whilst ensuring safe delivery. We use a home-made CO2 delivery system (Figure 1), similar to Kawasaki et al.6 Our sealed system consists of a gas cylinder, a 3-way stopcock, a CO2 bag and a 20-ml luer lock. The source of CO2 must be confirmed as a closed system circuit, free from air contamination. We fill the closed system with CO2 and purge the gas four times to minimize contamination. When the CO2 bag is sufficiently filled, we ensure we disconnect the CO2 cylinder to control volume injection and avoid excess delivery. We then discard approximately 50 ml from the system to reduce air contamination from the connective tubing. Prior to direct injection of CO2 contrast agent, we connect the syringe, CO2 bag and sheath via a 3-way stopcock. Our RDN technique is based on that described in the Symplicity hypertension (HTN)-2 trial and previous publications.2,4,7
Figure 1.
A home-made carbon dioxide delivery system: (A) a pressurized CO2 cylinder with filter, (B) a 50-ml syringe, attached to a one-way valve, and (C) a Merit disposal depot reservoir bag with removable scissor seal.
Each patient consented for the procedure and potential complications including abdominal pain, bleeding from puncture site, pseudoaneurysm, renal artery dissection, renal artery thrombosis and failure of procedure. The following procedures were conducted in accordance with the ethical standards of the responsible committee. Patient monitoring including oxygen saturation, electrocardiogram and blood pressure is necessary to identify acute deteriorations. Sedation and i.v. narcotics are routinely used. Initially, midazolam 5 mg i.v. and fentanyl 50 mcg i.v. are administered and subsequently titrated to requirement. Local anaesthetic, 1% lignocaine 10 ml, is injected into the right groin.
The patient is positioned to elevate the target site, ideally by 45o, to utilize gaseous reflux to reduce incomplete vasculature filling. The common femoral artery (CFA) is punctured using ultrasound guidance and a 6-Fr guide sheath introduced using a standard Seldinger technique. After the initial puncture, a 5-Fr pigtail catheter over a standard J-wire is introduced. Using the first lumbar vertebra as the central anatomical landmark, a digital subtraction aortogram is taken using CO2 as the contrast agent. To reduce air contamination, a small volume, approximately 5 ml, is initially expelled when the sheath is attached to the closed CO2 circuit. CO2 is slowly injected by hand via the delivery syringe. The rate of CO2 injection should be faster than the rate of blood flow to facilitate complete filling; we aim for an injection rate of 10–20 ml s−1. To allow for full elimination, injections of CO2 contrast are separated by 2–3 min. Theoretically, owing to rapid first-pass elimination in the lungs, there is no limit to the volume of CO2 used. However, 100 ml is reported as the maximum recommended amount.8 High-rate frames and image stacking improve the quality of the digital subtraction angiography as they reduce the effects of gas diffusion. The renal artery is accessed using a 5-Fr Cobra catheter™ (Boston Scientific, Malborough MA 01752). We perform a selective angiogram using CO2 to demarcate the renal artery, ostia, bifurcation and accessory arteries.
To manage coagulation, we administer 100 iU/kg of heparin. To prevent vasospasm, we administer an intra-arterial vasodilator, Isoket 100 mcg i.v., via the guide catheter. The Symplicity Flex™ catheter (Medtronic) is inserted at an initial position of 5 mm proximal to bifurcation of the renal artery at a posteroinferior angle. We have also used the new multielectrode 4-Fr Symplicity Spyral catheter (Medtronic) that allows for simultaneous ablations, thus reducing ablation time as only one burn cycle is required. The vessel resistance is checked to ensure it is >300 Ω. Additional analgesia, fentanyl 50 mcg i.v., is administered at this stage to pre-empt pain related to radiofrequency ablation delivery. The foot pedal is used for a 2-min burn cycle when radiofrequency ablation occurs via the Symplicity Flex catheter (Medtronic) tip electrode. The second burn is completed 5 mm proximal with a rotation of 45o. Repeating this burn cycle will treat the full circumference of the artery, effectively targeting the sympathetic nerve fibres. This helical burn cycle technique is continued up to 5 mm from the renal ostium at the superior surface of the renal artery. The above process is simplified when using the Symplicity Spyral catheter (Medtronic), as the helical shape simultaneously provides four-quadrant burns in a 1-min burn cycle. The Symplicity™ catheter (Medtronic) is removed and cleaned; the guide catheter is aspirated and discarded. A repeat selective right renal arteriogram is taken to ensure patency post-ablation. A selective left renal arteriogram is performed and the process is repeated for the left-hand side. Figure 2 demonstrates images obtained from the above protocol. We routinely and safely complete simultaneous bilateral RDN. Following successful completion of the procedure, a 6-Fr Angioseal™ St Jude Medical Minnetonka MN 55345-2126 USA vascular closure device is deployed.
Figure 2.
(a) Baseline selective left renal artery carbon dioxide (CO2) angiogram to assess dimensions of the main renal artery segment prior to renal sympathetic denervation (RDN); (b) post-RDN selective renal artery CO2 angiogram; (c) a Simplicity™ (Medtronic) device within the left renal artery during RDN procedure.
CONCLUSION
CO2 angiography in RDN is a relatively straightforward and feasible procedure, which could facilitate treatment of resistant hypertension in patients with renal insufficiency or iodinated contrast allergy. In our experience, images from CO2 angiography are of adequate quality to facilitate accurate RDN. We have found that the procedure is more efficient with the use of the multielectrode Symplicity Spyral system (Medtronic) and predict further technological advancements integrating CO2 angiography equipment will enhance the procedure significantly. We are conducting a clinical trial to investigate the clinical efficacy and safety of CO2 angiography for RDN in patients with moderate to severe renal impairment.
FUNDING
J Freedman and I Dasgupta have received an unrestricted research grant from Medtronic.
Contributor Information
Mary Renton, Email: mary.renton@nhs.net.
Mohammad A Hameed, Email: awaishameed@doctors.org.uk.
Indranil Dasgupta, Email: Indranil.dasgupta@heartofengland.nhs.uk.
Edward T D Hoey, Email: edward.hoey@heartofengland.nhs.uk.
Jonathan Freedman, Email: jsfreedman@doctors.org.uk.
Arul Ganeshan, Email: aganeshan@hotmail.com.
REFERENCES
- 1.Lloyd-Jones D, Adams R, Carnethon M, De Simone G, Ferguson TB, Flegal K, et al. Heart disease and stroke statistics—2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 2009; 119: e21–181. doi: https://doi.org/10.1161/CIRCULATIONAHA.108.191261 [DOI] [PubMed] [Google Scholar]
- 2.Krum H, Schlaich M, Whitbourn R, Sobotka PA, Sadowski J, Bartus K, et al. Catheter-based renal sympathetic denervation for resistant hypertension: a multicentre safety and proof-of-principle cohort study. Lancet 2009; 373: 1275–81. doi: https://doi.org/10.1016/S0140-6736(09)60566-3 [DOI] [PubMed] [Google Scholar]
- 3.Symplicity HTN-2 Investigators; Esler MD, Krum H, Sobotka PA, Schlaich MP, Schmieder RE, Böhm M. Renal sympathetic denervation in patients with treatment-resistant hypertension (The Symplicity HTN-2 Trial): a randomised controlled trial. Lancet 2010; 376: 1903–9. doi: https://doi.org/10.1016/S0140-6736(10)62039-9 [DOI] [PubMed] [Google Scholar]
- 4.Kawasaki D, Fujii K, Fukunaga M, Fukuda N, Masuyama T, Ohkubo N, et al. Safety and efficacy of carbon dioxide and intravascular ultrasound-guided stenting for renal artery stenosis in patients with chronic renal insufficiency. Angiology 2015; 66: 231–6. doi: https://doi.org/10.1177/0003319714524297 [DOI] [PubMed] [Google Scholar]
- 5.Fujihara M, Kawasaki D, Shintani Y, Fukunaga M, Nakama T, Koshida R, et al. Endovascular therapy by CO2 angiography to prevent contrast-induced nephropathy in patients with chronic kidney disease: a prospective multicenter trial of CO2 angiography registry. Catheter Cardiovasc Interv 2015; 85: 870–7. doi: https://doi.org/10.1002/ccd.25722 [DOI] [PubMed] [Google Scholar]
- 6.Kawasaki D, Fujii K, Fukunaga M, Masutani M, Nakata A, Masuyama T. Safety and efficacy of endovascular therapy with a simple homemade carbon dioxide delivery system in patients with ileofemoral artery diseases. Circ J 2012; 76: 1722–8. doi: https://doi.org/10.1253/circj.CJ-11-1382 [DOI] [PubMed] [Google Scholar]
- 7.Cho KJ. Carbon dioxide angiography: scientific principles and practice. Vasc Specialist Int 2015; 31: 67–80. doi: https://doi.org/10.5758/vsi.2015.31.3.67 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Kerns SR, Hawkins IF, Jr. Carbon dioxide digital subtraction angiography: expanding applications and technical evolution. AJR Am J Roentgenol 1995; 164: 735–41. doi: https://doi.org/10.2214/ajr.164.3.7863904 [DOI] [PubMed] [Google Scholar]