Table 3.

Air kerma rates in mGy s−1 delivered by the irradiator for different combinations of tube voltages (kVp) and tube currents (mA)

mA kVp
225 100 40
0.1 0.33 0.19 0.03
0.3 1.00 0.53 0.09
0.5 2.00 0.88 0.15
0.7 2.67 1.25 0.22
1 3.67 1.79 0.32
5 11.3 8.98 1.58
7 18.7 12.6 2.21
10 17.9 3.16
13 48.3
20 >tube max. power 35.9 6.31
30 >tube max. power 53.7 9.44

max., maximum.

Uncertainties on air kerma rates do not exceed 2.3% for the 225-kVp beams and 3.5% for the 100- and 40-kVp beams.

– : no measurement performed for these kVp-mA combinations.