DosiRat sensitivity and stem effect in pC Gy−1 for different beam qualities

Beam quality 225 kVp
0.3 mm Cu
100 kVp
2 mm Al
40 kVp
2 mm Al
Sensitivity (pC/Gy) 202.6 ± 4.7 126.4 ± 7.8 90.6 ± 5.7
Stem effect (pC/Gy) 0.19 ± 0.01 0.12 ± 0.01 0.12 ± 0.01
Stem effect contribution 0.10% 0.10% 0.13%

Al, aluminium; Cu, copper.

The stem effect proportion in DosiRat signal is also reported.