Table 2.
Study (chronological order) | Phantom ID | Test units | Voxel sizes (range) (mm) | FOVa (small, medium and large) | Evaluated image quality
parameters |
|||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Density values | Image uniformity | Presence of artefacts | Noise | Spatial resolution | Contrast | Geometric accuracy | Others | |||||
Araki et al14 (2004) | I | MercuRay (Hitachi Medical Systems America, Twinsburg, OH) | 0.1–0.376 | Small, medium and large | x | x | x | |||||
Marmulla et al15 (2005) | II | NewTom 9000 (QR srl, Verona, Italy) | 0.29 | nd | x | |||||||
Lagravere et al16 (2006) | III | NewTom QR-DVT 9000 (QR srl, Verona, Italy) | nd | nd | x | |||||||
Ballrick et al17 (2008) | IV | i-CAT (Imaging Sciences International, Hatfield, PA) | 0.2–0.4 | Small and medium | x | x | FOV size accuracy | |||||
Loubele et al18 (2008) | V | NewTom 3G (QR srl, Verona, Italy), Accuitomo 3D (J. Morita Mfg Corp., Kyoto, Japan), MercuRay (Hitachi Medical Systems America, Twinsburg, OH) | 0.13–0.38 | nd | x | x | x | |||||
Loubele et al38 (2008) | V | NewTom 3G (QR srl), i-CAT (Imaging Sciences International), MercuRay (Hitachi Medical Systems America), Accuitomo 3D (J. Morita Mfg Corp.) | 0.18–0.38 | Small, medium and large | x | |||||||
Bryant et al19 (2008) | VI | i-CAT (Imaging Sciences International) | 0.4 | Large | x | x | x | x | x | |||
Suomalainen et al20 (2009) | VII | 3D Accuitomo (J. Morita Mfg Corp.), Promax 3D (Planmeca, Helsinki, Finland), Scanora 3D (Soredex, Tuusula, Finland) | 0.12–0.36 | Small | x | x | x | |||||
Watanabe et al21 (2010) | VIII | 3D Accuitomo (J. Morita Mfg Corp.) | 0.125 | Small | x | |||||||
Vassileva and Stoyanov22 (2010) | IX | Iluma (IMTEC, Ardmore, OK) | 0.1–0.4 | Large | x | x | x | |||||
Pauwels et al23 (2011) | X | Scanora 3D (Soredex), Galileos (Sirona, Bensheim, Germany), i-CAT Classic (Imaging Sciences International), Iluma Elite (IMTEC), ProMax 3D (Planmeca), SkyView (MyRay, Imola, Italy), Veraviewepocs 3D (J. Morita Mfg Corp, Kyoto, Japan) | 0.13–0.35 | Small, medium and large | x | x | x | x | x | x | ||
Nackaerts et al24 (2011) | XI | Accuitomo 3D XYZ (J. Morita Mfg Corp.), Galileos Comfort (Sirona), Kodak 9000 3D (Kodak Dental Systems, Carestream Health, Rochester, NY), Picasso Duo (Vatech/E-WOO Technology, Seoul, Republic of Korea), Scanora 3D (Soredex) | 0.0765–0.289 | Small and medium | x | |||||||
Tsutsumi et al25 (2011) | XII | MercuRay 12 (Hitachi Medical Systems America) | 0.2–0.377 | Medium and large | x | |||||||
Panzarella et al26 (2011) | XIII | NewTom 3G (QR srl) and i-CAT (Imaging Sciences International) | 0.16–0.4 | Medium and large | x | |||||||
Watanabe et al9 (2011) | VIII | 3D Accuitomo (J. Morita Mfg Corp.) | 0.125 | Medium and large | x | |||||||
Reeves et al27 (2012) | XIV | Asahi Alphard 3030 (Belmont Takara, Kyoto, Japan) and Planmeca ProMax 3D (Planmeca, Helsinki, Finland) | 0.1–0.39 | Small, medium and large | x | |||||||
Pauwels et al39 (2012) | X | 3D Accuitomo 170 (J. Morita Mfg Corp.), 3D Accuitomo XYZ (J. Morita Mfg Corp.), Veraviewepocs 3D (J. Morita Mfg Corp.), Galileos Comfor (Sirona), i-CAT Next Generation (Imaging Sciences International), Kodak 9000 3D (Kodak Dental Systems, Carestream Health), Kodak 9500 (Kodak Dental Systems, Carestream Health), NewTom VGi (QR srl), Pax-Uni3D (Value Added Technologies, Yongin, Republic of Korea), Picasso Trio (Value Added Technologies, Yongin, Republic of Korea), ProMax 3D (Planmeca), Scanora 3D (Soredex) and SkyView (Cefla Dental Group, Imola, Italy) | 0.076–0.4 | Small, medium and large | x | x | ||||||
Thongvigitmanee et al28 (2013) | XV | DentiiScan (NECTEC and MTEC, Pathum Thani, Thailand) | 0.4 | Medium | x | |||||||
Ozaki et al29 (2013) | XVI | 3DXFPD8 (Hitachi Medical Systems America, Twinsburg, OH) and FineCube v. 12 (Yoshida Dental Mfg Co. Ltd, Tokyo, Japan) | 0.08–0.157 | Small | x | |||||||
Batista et al30 (2013) | XVII | Kodak 9000 (Kodak Dental Systems, Carestream Health), Kodak 9500 (Kodak Dental Systems, Carestream Health), and i-CAT Classic (Imaging Sciences International) | 0.076–1.0 | Small, medium and large | x | x | x | x | x | |||
Ludlow and Walker31 (2013) | XVIII | i-CAT Next Generation (Imaging Sciences International) | 0.125–0.4 | Small | x | x | x | x | Nyquist frequency | |||
Bamba et al40 (2013) | X | CS9300 3D (Kodak Dental Systems, Carestream Health, Rochester, NY), Accuitomo 80 (J. Morita Mfg Corp.) and Veraviewepocs 3Df (J. Morita Mfg Corp.) | 0.08–0,125 | Small | x | x | x | x | x | x | x | |
Torgersen et al10 (2014) | XIX | 3D Accuitomo 170 (J. Morita Mfg Corp.) | nd | Small | x | x | x | x | x | x | Nyquist frequency | |
Steiding et al32 (2014) | XX | Kavo 3D eXam (KaVo Dental GmbH, Biberach, Germany) | 0.3 | Medium | x | x | x | x | x | x | ||
Oliveira et al33 (2014) | XXI | Picasso-Trio (Vatech/E-WOO Technology, Seoul, Republic of Korea) | nd | Small | x | |||||||
Pauwels et al41 (2014) | X | 3D Accuitomo 170 (J. Morita Mfg Corp.) | 0.16 | Small and medium | x | |||||||
Dillenseger et al34 (2015) | XXII | Newtom 5G (QR srl, Verona, Italy) | 0.075–0.25 | Small and medium | x | x | x | x | x | |||
Plachtovics et al35 (2015) | XXIII | iCAT Classic (Imaging Sciences International) | 0.2 | Small | x | x | ||||||
Pauwels et al36 (2015) | XXIV | 3D Accuitomo 170 (J. Morita Mfg Corp.), Cranex 3D (Soredex, Tuusula, Finland), Scanora 3D (Soredex), and Galileos Comfort (Sirona) | 0.125–0.35 | Small, medium and large | x | |||||||
Kosalagood et al42 (2015) | X | 3D Accuitomo 170 (J. Morita Mfg Corp.), Galileos Comfort (Sirona), Scanora 3D (Soredex), CB MercuRay (Hitachi Medical Systems America), i-CAT Next Gen (Imaging Sciences International), Kodak CS 9300 (Kodak Dental Systems, Carestream Health), WhiteFox (Acteon Group, Mérignac, France), Dentiiscan (NECTEC and MTEC, Pathum Thani, Thailand) | 0.25–0.4 | Small, medium and large | x | |||||||
Steiding et al12 (2015) | XX | KaVo 3D eXam (KaVo Dental GmbH) | 0.3 | Medium | x | x | x | x | x | |||
Abouei et al43 (2015) | X | Carestream 9300 (Kodak Dental Systems, Carestream Health, Rochester, NY) | 0.18–0.25 | Small and medium | x | x | x | x | x | x | ||
Ali et al44 (2015) | VIII | Scanora 3D (Soredex) and 3D Accuitomo 80 (J. Morita Mfg Corp.) | 0.08 | Small | x | x | ||||||
Choi et al45 (2015) | X | Dinnova3 (HDXwill Inc., Seoul, Republic of Korea) | 0.3 | Large | x | x | x | |||||
Pauwels et al46 (2016) | X | 3D Accuitomo 170 (J. Morita Mfg Corp.) | nd | Small and medium | x | |||||||
Taylor et al47 (2016) | X | 3D Accuitomo 170 (J. Morita Mfg Corp.) | nd | Medium | x | x | x | |||||
Elkhateeb et al37 (2016) | XXV | CS 9300 PREMIUM (Kodak Dental Systems, Carestream Health, Rochester, NY) | 0.18–0.5 | Small and medium | x | x | x | x |
FOV, field of view; nd, not declared.
Small FOV = Height ≤ 80 mm, Medium FOV = Height > 80 ≤ 130 mm, Large FOV = Height > 130 mm.