Table 1. Operating parameters for REE determination using ICP-OES and ICP-MS.

Instrument settings ICP-OES ICP-MS
Rf power 1000 W 1100 W
Plasma gas (Ar) flow rate 15 L min-1 15 L min-1
Auxiliary gas (Ar) flow rate 1.5 L min-1 1.2 L min-1
Nebulizer gas flow rate 0.6 L min-1 0.8 L min-1
Nebulizer Concentric Concentric
Dwell time - 50 ms
Pump speed 15 rpm -
Nebulizer pressure 200KPa -
Repetition views 2 3
Analytical wavelength (nm) Ce 418.659, Dy 353.170, -
Er 337.275, Eu 381.967,
Gd 335.047, Ho 345.600
La 408.672, Lu 261.542,
Nd 401.225, Pr 417.939,
Sm 359.259, Tb 350.917,
Tm 313.126, Yb 328.937,
Y 324.228, Sc 361.383