Table 8.

Between-run precision in dog VS, established from three independent runs on different days, using two different lots of Enzolifesciences EIA kits.

Run 1 Run 2 Run 3 Between-run mean concentration (pg/ml) Between-run precision (%CV) Precision acceptance criteria (%)
Mean concentration (pg/ml) Within-run precision (%CV) Precision acceptance criteria (%) Mean concentration (pg/ml) Within-run precision (%CV) Precision acceptance criteria (%) Mean concentration (pg/ml) Within-run precision (%CV) Precision acceptance criteria (%)
VS A 26.1 3.7 ≤25 39.7 31.9 ≤25 52.7 23.0 ≤20 39.5 33.7 ≤25
VS B 73.5 10.5 ≤20 67.3 23.8 ≤20 79.9 27.7 ≤20 73.5 8.6 ≤20
VS C 132.8 20.0 ≤20 124.7 24.0 ≤20 125.9 14.3 ≤20 127.8 3.4 ≤20

Values out of the acceptance range (i.e., if CV >20%, or >25% at the interval between 1x and 3xLLOQ) are indicated in bold.