Skip to main content
. 2017 Aug 24;2017:1202459. doi: 10.1155/2017/1202459

Table 1.

Validation parameters of the LC-MS/MS method.

Analyte r LLOQ (ng/g) Recovery (%) Intraprecision (RSD %) Interprecision (RSD %)
DA 0.9969 2.0 94.6 ± 8.7 10.3 12.4
DOPAC 0.9982 72.0 93.4 ± 6.6 8.9 7.5
HVA 0.9977 50.0 95.3 ± 7.4 7.7 11.3

DA: dopamine; DOPAC: 3,4-dihydroxyphenylacetic acid; HVA: homovanillic acid.