Table 3.

Results of the Patient Study.a

Difference Vertical (mm) Longitudinal (mm) Lateral (mm) Rotation (°) Roll (°) Pitch (°)
Minimum 0.02 0.03 0.01 0.15 0 0
Maximum 0.64 0.79 0.62 0.72 0.48 0.47
Mean 0.20 0.30 0.21 0.40 0.17 0.20
Standard deviation 0.16 0.18 0.18 0.16 0.13 0.14

Abbreviation: CBCT, cone-beam computed tomography.

aResidual error differences (absolute values) in translational and rotational directions, between ExacTrac X-ray imaging registration and TrueBeam CBCT registration among 34 patient studies.