Skip to main content
. 2016 Jan 20;16(1):45–56. doi: 10.1177/1533034615626011

Table 2.

Gafchromic EBT-2 Film Measurement of FWHM and 80% to 20% Penumbra at Different Depths for Various Collimators.

Collimator At Isocenter At 18.6 mm Depth At 37.2 mm Depth
0.5 mm diameter FWHM (mm) 0.47 0.54 0.61
Penumbra (mm) 0.2 ± 0.1 0.3 ± 0.1 0.3 ± 0.1
1 mm diameter FWHM (mm) 1.14 1.18 1.2
Penumbra (mm) 0.3 ± 0.1 0.4 ± 0.1 0.5 ± 0.1
2 × 2 mm2 FWHM (mm) 2.46 2.57 2.63
Penumbra (mm) 0.5 ± 0.1 0.58 ± 0.1 0.5 ± 0.1
4 × 4 mm2 FWHM (mm) 4.48 4.81 5.01
Penumbra (mm) 0.5 ± 0.1 0.5 ± 0.1 0.6 ± 0.1