Skip to main content
. 2016 Nov 22;3:866–869. doi: 10.1016/j.toxrep.2016.11.001

Table 2.

Working conditions for determination of concentration of some heavy metals using Atomic absorption spectrophotometer.

Metals Wavelength (nm) Slit width (nm) Detection limit (mg/l) Lamp current (mA) Linear range (mg/l) Flame type (colour)
Mn 279.5 0.7 0.03 10 2.50 A-A, lean/blue
Cd 228.80 0.7 0.01 2 2.00 A-A, lean/blue
Cr 357.90 0.7 0.04 2 5.00 A-A, rich/yellow
Cu 324.80 0.7 0.005 1.5 5.00 A-A, lean/blue
Pb 283.30 0.7 0.08 5 20.00 A-A, lean/blue
Ni 232.00 0.2 0.05 2.5 4.00 A-A, lean/blue