Skip to main content
. 2017 Sep 20;7(3):39. doi: 10.3390/bios7030039

Figure 2.

Figure 2

Figure 2

Diagnostic accuracy of the paper-based SCA screening test kit. (a) Aggregate confusion matrix for screening of blood samples (n = 185) with characteristic HbS concentrations via visual interpretation (inexperienced users; n = 3) of the blood stains produced on paper by the metabisulfite (MS) formulation of the test (185 samples × 3 users = 555 total scores). (b) Aggregate confusion matrix for the screening of blood samples (n = 143) with characteristic HbS concentrations via visual interpretation (inexperienced users; n = 3) of the blood stains produced on paper by the hydrosulfite (HS) formulation of the test (143 samples × 3 users = 429 total scores). Rows correspond to characteristic genotypes (based on HbS concentration measured using conventional high-performance liquid chromatography—HPLC) and columns correspond to predicted genotypes (diagnosed by the paper-based test). Shaded cells contain the numbers of correct diagnoses.