Table 2.

Comparison of the CT number of three ROIs of the anthropomorphic pelvis phantom. For the proposed method, three blockers were used, they have the same strip width of 20 pixels (B20) and different gap width 20, 40, 60 pixels (G20, G40, G60), three moving speeds were used (8, 15, 23 pixels per projection)

Blocker Moving speed (pixels per projection) ROI1 mean/stda ROI2 mean/std ROI3 mean/std RMSE e
G20 8 102/44 106/45 54/12 35
G20 15 89/40 98/45 51/13 30
G20 23 100/43 104/46 55/12 34
G40 8 74/39 96/56 42/8 26
G40 15 72/43 93/50 40/11 25
G40 23 73/38 103/55 35/5 25
G60 8 76/41 102/43 31/7 22
G60 15 76/40 95/44 12/6 13
G60 23 64/30 94/47 21/7 19
Varian scatter correction images 62/32 75/17 5/9 15
No scatter correction images ‐253/34 ‐177/52 ‐261/10 290
Simulated scatter free images 88/13 79/12 3/5
a

Standard Variation.