Table 1.

Performance summary and comparison.

Reference This Work ISSCC ‘12 [14] TBioCAS 2012 [15] Sensors 2014 [17] JSSC‘16 [18] JSSC’12 [20] TMTT‘13 [19] ISSCC ‘16 [34]
CMOS technology (nm) 180 250 250 350 350 90 90 180
Architecture Phase-based Time-based Time-based Time-based Time-based Freq.-based Freq.-based Freq. ratio
Q factor 130 280 450 386 376 N/A N/A N/A
Resonant frequency (Hz) 898 M 1.98 M 2.17 M 535.8 k 592 k 7~9 G 10.4 G 45 M
Frequency resolution (Hz) 733 53 5 26.8 17.6 2 M 156 k N/A
Relative freq. resolution 1 (ppm) 0.82 26.77 2.3 50 29.8 >222 15 0.00028
Power cons. (mW) 7 1.35 1.35 0.1 0.06 16.5 22 19
Measurement time (ms) 0.48 5000 10,000 6 1.1 0.9 25 3.85

1 The relative frequency resolution is defined as the frequency resolution divided by the device resonance frequency.