Published in final edited form as: J Chromatogr A. 2017 Aug 26;1520:75–82. doi: 10.1016/j.chroma.2017.08.050

Table 4.

The Overall Prediction Error (s) of 13 Test Compounds When Using Offset Gradients

Overall Prediction Error (s)
Flow Rate 0.2 mL/min 0.4 mL/min 0.8 mL/min
Gradient Slope
3.5%/min g ± 7.2e (2.25% f) ± 3.3 (1.63%) ± 3.0 (2.73%)
1.75%/min h ± 6.3 (1.63%) ± 4.2 (1.65%) ± 3.9 (2.67%)
0.875%/min i ± 7.0 (1.56%) ± 4.4 (1.54%) ± 4.4 (2.62%)
e

Overall prediction error is defined as the RMS (root mean square) error of the test compounds. RMS(1n(x12+x22+x32+xn2)), x means the prediction error of each test compound

f

Predictionerror(%)=(1n(y12+y22+y32+yn2)), y means the prediction error (%) of each test compound

g

3.5%/min is the gradient slope of 3.5% mobile phase B/min, mobile phase B: water with 20 mM ammonium acetate (pH 4.7)

h

1.75%/min is the gradient slope of 1.75% mobile phase B/min, mobile phase B: water with 20 mM ammonium acetate (pH 4.7)

i

0.875%/min is the gradient slope of 0.875% mobile phase B/min, mobile phase B: water with 20 mM ammonium acetate (pH 4.7)