Table 1.

Range, accuracy, intermediate precision, and specificity of relative binding assay. LLOQ, lower limit of quantitation; LQC, low‐quality control; MQC, middle‐quality control; HQC, high‐quality control; ULOQ, upper limit of quantitation

Fcγ receptor Range (μm) Test sample Nominal value (IgG; (μm) Accuracy (%) Intermediate precision (%) Specificity: % binding of IgA
FcγRIIIa 0.104–3.33 LLOQ 0.1 96.6 7.75
LQC 0.2 97.6 5.59 0.5
MQC 0.9 106.0 5.33 1.8
HQC 2.7 84.4 8.08
ULOQ 3.1 93.8 5.13
FcγRIIIb 0.832–26.67 LLOQ 0.9 110.6 4.48
LQC 3.1 87.1 2.70 9.9
MQC 6.7 91.5 5.79 13.2
HQC 20.0 104.4 7.12
ULOQ 22.2 104.3 11.99
FcγRIIa 0.104–3.33 LLOQ 0.1 111.4 7.99
LQC 0.2 90.6 4.09 1.4
MQC 0.9 92.0 7.50 3.3
HQC 2.7 101.5 12.11
ULOQ 3.1 101.4 7.33
FcγRIIb 0.832–26.67 LLOQ 0.9 106.2 5.93
LQC 3.1 90.3 3.64 14.3
MQC 6.7 96.1 4.45 27.9
HQC 20.0 101.5 6.81
ULOQa 22.2 101.0 6.83
a

One sample was excluded due to an air bubble in the injection; n = 17. All other results are based on n = 18.