Table 1.

Imaging protocols depending on patient position and renal function.

Protocol A n = 5 Protocol B n = 8 Protocol C n = 10 Protocol D n = 12
Patients renal function Partial renal function Partial renal function Anuria Anuria
Central drain evaluable no yes no yes
Extremity positioning Extremity over head Extremity 90° angulated beside the thorax Extremity over head Extremity 90° angulated beside the thorax
Patient position Abdominal position Supine position Abdominal position Supine position
Scan parameter
 Tube voltage [kV] 70 kV 80 kV 70 kV 80 kV
 Effective current-time product [mAs] 150/180 mAs 180 mAs 150/180 mAs 180 mAs
 Dose length product [mGy*cm] 716 ± 58 1246 ± 76 734 ± 63 1223 ± 82
 Effectiv radiation dose [mSv] 0.36 ± 0.03 17.4 ± 1.1 0.37 ± 0.03 17.1 ± 1.2
 Scan range [mm] 265–340 454–630 265–340 454–630
Scan cycles
 Temporal resolution 2 sec 6 6
 Temporal resolution 4 sec 4 4
 Temporal resolution 6 sec 2 2
 Temporal resolution 2.5 sec 6 6
 Temporal resolution 5 sec 4 4
 Temporal resolution 7.5 sec 2 2
 Total scan time [sec] 40 sec 50 sec 40 sec 50 sec
Contrast media (iomeprol 400 mg iodine/ml)
 Amount 18 g/45 ml 18 g/45 ml 32 g/80 ml 32 g/80 ml
 Flow rate [ml/sec] 5 ml/sec 5 ml/sec 5 ml/sec 5 ml/sec

Notes kV = kilo voltage.