. 2017 Sep 25;55(10):3037–3045. doi: 10.1128/JCM.00806-17

TABLE 4.

Discrepant resolution by repeating illumigene MP test and in-house qPCR to detect Plasmodium spp.

Study IDa Resultb
Microscopy illumigene malaria Plus test
In-house qPCR Final
Original Repeat
N118 NEG POS Not donee NEG FP
N55 NEG POS NEG NEG TN
N34 NEG POS NEG NEG TN
S32 NEG POS Not donec POS (P. falciparum) TP
S62 NEG POS Not donec POS (P. falciparum) TP
N17 P. vivax Invalid POS Not doned TP
N23 P. vivax Invalid POS Not doned TP
N57 NEG Invalid NEG Not doned TN
N59 NEG Invalid NEG Not doned TN
N75 NEG Invalid NEG Not doned TN
N80 NEG Invalid Invalid NEG Invalid
N93 NEG Invalid POS POS (P. falciparum) TP
S82 P. ovale Invalid POS Not doned TP
a

ID, identifier.

b

NEG, negative; POS, positive; TP, true positive; FP, false positive; TN, true negative.

c

illumigene test not repeated if real-time PCR agreed with original MP kit result.

d

qPCR was not done if illumigene MP repeat result was concordant with microscopy.

e

Insufficient sample volume to repeat test.