TABLE 4.
Study IDa | Resultb |
||||
---|---|---|---|---|---|
Microscopy |
illumigene malaria Plus test |
In-house qPCR | Final | ||
Original | Repeat | ||||
N118 | NEG | POS | Not donee | NEG | FP |
N55 | NEG | POS | NEG | NEG | TN |
N34 | NEG | POS | NEG | NEG | TN |
S32 | NEG | POS | Not donec | POS (P. falciparum) | TP |
S62 | NEG | POS | Not donec | POS (P. falciparum) | TP |
N17 | P. vivax | Invalid | POS | Not doned | TP |
N23 | P. vivax | Invalid | POS | Not doned | TP |
N57 | NEG | Invalid | NEG | Not doned | TN |
N59 | NEG | Invalid | NEG | Not doned | TN |
N75 | NEG | Invalid | NEG | Not doned | TN |
N80 | NEG | Invalid | Invalid | NEG | Invalid |
N93 | NEG | Invalid | POS | POS (P. falciparum) | TP |
S82 | P. ovale | Invalid | POS | Not doned | TP |
ID, identifier.
NEG, negative; POS, positive; TP, true positive; FP, false positive; TN, true negative.
illumigene test not repeated if real-time PCR agreed with original MP kit result.
qPCR was not done if illumigene MP repeat result was concordant with microscopy.
Insufficient sample volume to repeat test.