Table 2.

Summary of qualification results for linearity, accuracy, Limit of Quantitation (LOQ), and Limit of Detection (LOD).

  Linearity and Range
    Accuracy
  R2 Range LOQ, from calibration curve, 10 (SD/S) LOD, from calibration curve, 3.3 (SD/S) Level % Recovery % CV
CDR isomerization (%) 0.9995 4.1 – 65.7 3.3 1.1 Low 107.0 1.7
          Middle 99.0 0.4
          High 100.0 0.8
CDR deamidation (%) 0.9981 3.2 – 10.7 1.2 0.4 Low 106.0 0.3
          Middle 102.0 0.9
          High 103.0 0.5
PENNY deamidation (%) 0.9996 2.5 – 50.1 2.6 0.8 Low 103.0 1.7
          Middle 101.0 2.9
          High 99.0 0.9
Met-252 oxidation (%) 0.9977 3.1 – 46.2 4.8 1.6 Low 104.0 3.9
          Middle 96.0 2.4
          High 100.0 0.0
HL trisulfide bond (%) 0.9970 2.0 – 11.1 1.6 0.5 Low 96.0 0.6
          Middle 105.0 4.5
          High 103.0 8.8
HL thioether bond (%) 0.9979 0.5 – 13.6 1.4 0.5 Low 96.0 1.1
          Middle 106.0 0.1
          High 102.0 2.6
Cys-214 free thiol (%) 0.9936 0 – 3.1 0.6 0.2 Low 160.0 3.6
          Middle 99.0 1.3
          High 104.0 6.0
Afucosylation (%) 0.9917 8.3 – 31.4 6.7 2.2 Low 99.0 0.9
          Middle 98.0 0.3
          High 101.0 0.4