Skip to main content
. 2016 Aug 26;41(4):487–495. doi: 10.1016/j.jgr.2016.08.008

Table 4.

Accuracy of UPLC method for the determination of reference compounds

Analyte Original
(mg)
Spiked
(mg)
Found
(mg)
Recovery
(%)
RSD
(%)
Rg1 1.09 0.88 1.96 98.99 2.80
1.1 2.24 105.37 0.36
1.32 2.34 94.80 2.94
Re 4.19 1.2 5.48 106.83 2.85
1.5 5.69 99.53 0.85
1.8 6.01 101.05 2.62
Rb1 0.58 0.4 0.98 98.94 4.52
0.5 1.04 92.82 1.24
0.6 1.19 101.99 0.42
Rc 1.02 0.64 1.68 104.44 0.96
0.8 1.82 100.13 0.33
1.06 1.98 90.64 1.74
Rb2 0.70 0.44 1.15 100.80 0.90
0.55 1.24 97.98 0.57
0.66 1.34 97.27 1.24
Rb3 0.20 0.16 0.33 84.17 6.46
0.20 0.36 82.42 1.26
0.24 0.42 91.73 2.84
Rd 2 1.60 3.66 104.05 1.15
2.00 3.95 97.18 0.62
2.40 4.32 96.84 1.49

RSD, relative standard deviation; UPLC, . ultraperformance liquid chromatography