Skip to main content
. 2017 Oct 5;7:12727. doi: 10.1038/s41598-017-13097-6

Table 2.

R.m.s. resultant distortion values, drms (mm), for three categories of patient: prostate DW image volumes radiologically assessed as having ‘no distortion’ or ‘some distortion’ or patients with hip prostheses incurring the possibility of ‘severe image warping’ due to susceptibility effects.

Distortion (mm)
No distortion Distortion Distortion (hip patients)
Study drms s.e.m. Study drms s.e.m. Study drms s.e.m.
1 3.456 0.005 11 5.489 0.005 21 6.792 0.009
2 1.936 0.007 12 5.214 0.007 22 12.419 0.008
3 1.864 0.005 13 3.684 0.007 23 8.624 0.008
4 1.635 0.006 14 6.415 0.006 24 4.220 0.009
5 2.097 0.007 15 4.449 0.007 25 4.645 0.006
6 1.821 0.007 16 4.015 0.005 26 14.213 0.006
7 1.926 0.006 17 3.693 0.005 27 16.489 0.009
8 1.206 0.006 18 2.810 0.005 28 4.612 0.009
9 1.750 0.005 19 4.869 0.006 29 4.615 0.008
10 1.734 0.006 20 3.388 0.006 30 4.207 0.006
mean (sd) 1.943 (0.582) mean (sd) 4.402 (1.098) mean (sd) 8.083 (4.653)