Published in final edited form as: J Artif Organs. 2017 Apr 8;20(4):381–385. doi: 10.1007/s10047-017-0958-5

Table 1.

Hemodynamics with Pump Speed Modulation

Speed modulation
[±%]
Speed (mean)
[rpm]
Pump flow (mean)
[L/min]
AoP (mean)
[mm Hg]
AoP (pulse)
[mm Hg]
AoP (dP/dt)
[mm Hg/sec]
Pulsatility Index EEP
[mm Hg]
SHE
[ergs/cm]
PAP (mean)
[mm Hg]
PAP (pulse)
[mm Hg]
0 3,060 9.4 ± 1.1 91.6 ± 6.4 5.6 ± 2.2 67 ± 23 0.18 ± 0.06 91.6 ± 6.4 9 ± 19 32.5 ± 7.7 9.6 ± 3.7
15 3,010 9.8 ± 1.1 89.9 ± 4.2 12.0 ± 4.1* 91 ± 27 1.05 ± 0.14 90.9 ± 4.5 1,346 ± 479* 32.3 ± 7.9 12.3 ± 0.5
25 3,020 10.4 ± 1.5 88.9 ± 4.6 15.9 ± 5.3* 121 ± 28* 1.51 ± 0.20 91.5 ± 5.7 3,381 ± 1,477* 31.8 ± 8.5 15.7 ± 0.8*

AoP, systemic arterial pressure (obtained from a catheter within the carotid artery); dP/dt, the first derivative of the pressure; EEP, energy-equivalent pressure; SHE, surplus hemodynamic energy; PAP, pulmonary arterial pressure.

*

p < 0.05 compared with 0% speed modulation.

p < 0.01 compared with 0% speed modulation.