Table 1.

Participants encountered three standardized problem scenarios during the final assessment

Problem scenario Cause Symptoms Timing Correct steps
(1) Insufflator malfunction Gas tank closed upon start - Alarm insufflator (auditory) From start - Check insufflator
- Check gas tank
- Loss of pneumoperitoneum - Check gas tubes
- Insufflator gas bar empty - Check trocar position and valves
(2) Electro-surgery malfunction Patient grounding plate not fit - Alarm (auditory) From start electrocoagulation task (±3 min) - Check display electrosurgical unit
- Check cables
- Electrocoagulation failure - Check patient grounding plate
(3) Pulse saturation change Pulse oximeter malfunction - Auditory pulse signal fails to appear Simultaneous with scenario #2 - Check anesthesia monitor
- Flat line anesthesia monitor - Check pulse oximeter