Table 1.

Frequency of bacteria samples identified by VITEK 2, Phoenix, MALDI-MS and MALDI-BD according to identification score

Characteristic n
VITEK 2—Quality of identification
 Excellent 14
 Very good 4
 Good 2
 Low discrimination 2
MALDI-MS—Identification score
 Reliable genus and probable species (99.9% to 90%) 7
 Identification of probable genus (89.9% to 85%) 3
 Unreliable identification (<85%) 6
 Not identified 6
MALDI-BD—Identification score II
 Highly probable species 3
 Secure genus and probable species identification 10
 Probable genus 6
 Unreliable 3
Phoenix—Quality of identification
 100% to 96% 20
 95% to 93% 1
 92% to 90% 1
Total 22

MALDI-BD, MALDI-TOF BioTyper (Becton Dickinson (BD), San Diego, CA, USA); MALDI-MS, MALDI-TOF VITEK MS (bioMérieux, Marcy l’Étoile, France); MALDI-TOF, matrix-assisted laser desorption/ionization time-of-flight mass spectrometry; Phoenix (BD); VITEK 2, VITEK 2 automated phenotypic method (bioMérieux).